Effect of Bone Cement Viscosity and Set Time on Mantle Area in Total Knee Arthroplasty
Michael Kopec, BS, Joseph C. Milbrandt, PhD, Nick Kohut, BS, Brian Kern, MD, and D. Gordon Allan, MD, FRCS(C)
To assess the impact of bone cement viscosity on total knee arthroplasty, we compared 1 high-viscosity and 2 medium-viscosity cements with respect to mantle area and zone-specific intrusion depths into the tibial plateau. We analyzed postoperative radiographs to determine penetration area and depth in 72 consecutive patients (79 knees) in whom DePuy II (n = 11), Endurance (n = 34), or Simplex-P (n = 34) cement was used.
Penetration into the tibial plateau (anteroposterior zones 1-4) was significantly reduced with use of the high-viscosity DePuy II cement but did not differ significantly between the 2 medium-viscosity cements, Endurance and Simplex-P. Surgical and tourniquet times were significantly decreased with the quicker setting DePuy II cement.
Given these findings, additional studies are warranted to assess the long-term impact of the lower intrusion depths found with DePuy II cement. Such differences in cement penetration could jeopardize long-term fixation and lead to higher long-term device failure rates.