Low-Permeability Stent Reduces AAA Volume, Diameter at One Year


 

WASHINGTON — The low-permeability version of the Gore Excluder stent graft appears to reduce the volume and diameter of abdominal aortic aneurysms at 1 year significantly more than did the older, more permeable version of the stent, according to the results of a prospective study comparing the two models.

The high rate of failure with the original Gore Excluder stent graft, in which the mean maximum abdominal aortic aneurysm (AAA) diameter increased by at least 5 mm in 36% of patients after 5 years, prompted the stent's manufacturer, W.L. Gore & Associates Inc., to add a new low-permeability layer between the device's expanded polytetrafluoroethylene layer and its reinforcing membrane, Dr. Manish Mehta said at the annual meeting of the Eastern Vascular Society.

The Food and Drug Administration approved the new device in July 2004.

In a series of 428 patients who received either the new or the old Excluder for AAA treatment during 2001–2005 at the Institute for Vascular Health and Disease at Albany (N.Y.) Medical College, Dr. Mehta and his colleagues compared the outcomes of 214 patients who had CT angiography imaging results available 1 year after the procedure.

The CT angiograms revealed that 114 consecutive patients with the low-permeability stent had significantly greater declines in mean AAA volume than 100 consecutive patients with the original type of stent (163 mL to 141 mL vs. 156 mL to 160 mL). The percentage of patients with more than a 5% drop in AAA volume also was significantly greater in patients with the new stent (52%) than in those with the old stent (6%), reported Dr. Mehta, a vascular surgeon at the institute.

Compared with recipients of the old stent graft, significantly fewer patients with the new device had more than a 5% increase in AAA volume (12% vs. 2.6%).

A significantly greater percentage of patients with the low-permeability stent had at least a 5-mm decrease in maximum AAA diameter (25% vs. 8%). The mean preoperative maximum AAA diameter was similar among the patients with the new (5.4 cm) or old (5.3 cm) Excluder.

Recommended Reading

Early, Aggressive Statin Use Lowers Post-ACS Mortality
MDedge Cardiology
Five Types of MI Will Make Up New Definition
MDedge Cardiology
Men With Gout Have 26% Higher Risk of Acute Myocardial Infarction
MDedge Cardiology
Chronic Total Occlusion Success Rates Are Stalled
MDedge Cardiology
Strokes After Carotid Stenting Linked to Preventable Errors
MDedge Cardiology
Cryoablation in Valve Repair Halts Atrial Fib in 80%
MDedge Cardiology
Combined Monitoring Helps Avoid Neurologic Complications
MDedge Cardiology
Less Restenosis in New Everolimus-Eluting Stent in SPIRIT II
MDedge Cardiology
Post-MI Deaths Down With 'New' Sulfonylureas
MDedge Cardiology
Speedier ECGs Don't Affect Time To Reperfusion
MDedge Cardiology